Automatic
Steriliser / Autoclave Validation Tool
A
quick and easy method of validating Bench-Top autoclaves to
meet the requirements of Health Technical Memorandum 01-05:
‘Decontamination in primary care dental practices’
for best practice
|
-
Maintain Best Practice
- Speed
up Validations, need to stand over!
- Independently
verified
- See
performance at a glance
- Create
permanent inviolate record
- Re-use
up to 300 times
- Confirmation
to EN13060 & HTM 01-05.
Signatrol
ltd. has introduced a new quick response, low cost temperature
data logger specifically designed to meet the daily requirements
of bench-top autoclave validation.
The
kit comprises a SL55T-A data-logger, an anodised aluminium
enclosure and a stainless steel metal frame. |
|
The data logger is configured from a PC running the bespoke
TempIT software and is then inserted in the enclosure. The
whole assembly is then placed inside the autoclave and a validation
cycle performed. At the end of the cycle the logger is retrieved
and connected to the PC where the temperature vs time record
can be immediately displayed both graphically and in tabular
format. The readings are stored in encrypted form that cannot
be altered by ordinary means and therefore forms a permanent
inviolate record. Markers on the graph can easily be positioned
at the start and end of the sterilisation temperature (normally
134°C) and the time over this temperature can easily be
read off and the software will automatically indicate if the
measured cycle passes or fails the test.
What
you get in the Kit
- SL55T-A:
Data-logger with calibration certificate traceable to National
Standards by reference to equipment calibrated by a UKAS
certified laboratory'
- SL50-INTERFACE_USB:
Cable to connect the logger to the USB port of a PC computer
- SL50-ACC10:
Special steam resistant fast response enclosure
- SL50-ACC11:
Support Frame to ensure the logger is suspended clear of
any objects that might effect the cycle response
- TempIT-Pro
configuration and analysis software
|
| Specifications |
|
| Power |
Internal
Battery |
| Battery
Life |
|
| Temperature
Range |
+15°C to +140°C |
| Accuracy |
SL55T-A:
±0.5°C from +80°C to +140°C
|
| Resolution |
11bit - 0.07°C |
| Number
of Readings |
4096 high res mode |
| Response
Time |
63%
step change - 33 seconds in steam |
| IP
Rating |
IP68 |
| Sample
Rate |
|
| Software
Alarms |
Yes |
| Warranty |
1
Year |
| Dimensions
|
17mm
Diameter x 6mm High |
| Download
Software |
|
| Data
Logger Datasheet |
|
| Enclosure
Datasheet |
|
| TempIT
Software |
|
Part
Number |
Description |
Price
(Ex VAT) |
Buy |
HE-55A-KIT-1 |
Validation
kit containing one data logger, holding frame, USB interface
and software
|