Temperature

SL55T-D - Autoclave Validation Data Logger
£ 45.00 ex vat

Autoclave Validation Data Logger
(click to enlarge)

The SL55T-D is a replacement data logger for the HE-55D-KIT-1 autoclave validation system. It has the ability to record up to 8,000 temperature points over the range +15°C to +140°C. The SL55T-D functions the same as the SL55T but with an increased accuracy over the range +80°C to +140°C. The button data loggers themselves are very small (about the same size as a watch battery) which enables them to be inserted into small items and packages.

The SL50R Button Reader™ interfaces to the USB port of a PC and is totally compatible with the user friendly, TempIT-Lite operating software which enables the logger to be set up and the data read and displayed in graphical format. Data can be stored and retrieved at a later time. The software can be up-graded to TempIT-Pro which provides various additional features such as data table view, export data to spreadsheet, email graphs, and automated MKT, F0, A0 and PU calculations etc.

 Please note: If ordering online, there will be a slight delay in despatching the items due to the additional calibration procedure required to convert the 'standard' SL55T to the SL55T-D


Qty.  
Available
Tell a Colleague about this product.

Product Specifications



Specifications

SL55T-D (Improved Accuracy)

Power
Internal Battery
Battery Life

300 Hours when used above 50°C
*Low resolution mode
Battery Life Calculator Available Here

Temperature Range
+15°C to +140°C
Accuracy

±0.5°C from +80°C to +140°C
±1.0°C from +15°C to +80°C

Resolution

8bit - low res mode (0.5°C)
11bit - Hi res mode (0.07°C)

Number of Readings
8192 low res mode
4096 high res mode
Response Time
63% step change - 90 seconds in circulating air
10% to 90% step change - 180 seconds in circulating air
IP Rating
IP55
'Protected against ingress of dust and protected against low-pressure jets of water'
Please see waterproof enclosures if the product is to be submerged in liquid
Delayed Start
Yes, Up to 1 Year
1 minute resolution
Sample Rate

1 second to 24 hours

1 second to 1 minute in 5 second steps
1 minute to 1 hour in 1 minute steps
1 hour to 24 hours in 1 hour steps

Software Alarms
Yes
Intrinsic Safety Meets UL#913 (4th Edit.); Intrinsically Safe Apparatus, approval under Entity Concept for use in Class I, Division 1, Group A, B, C, and D Locations
Material
305 Stainless Steel
Warranty
1 Year
Dimensions
17mm Diameter x 6mm High
Download Software

 
Data Logger Datasheet
Enclosure Datasheet
View Independent Verification Report
 



SHOPPING CART
Your Basket: 0 Item(s)
Checkout
 
 
Related Products

Verifier

Verifier
Verifier On CD or FREE Download
Verifier application for SL55T-D dataloggers.Allows data logger configuration, downlo...
Add to Cart
£15.00 ex vat.

HE-55D-KIT-1

HE-55D-KIT-1
Autoclave Testing Kit
Autoclave testing kit containing one data logger, holding frame, USB interface and so...
Add to Cart
£130.00 ex vat.

Payment Method

Payment Method