SL7104 Four Channel Data Logger


SL7104 Four Channel Data Logger

The SL7104 data logger has a very flexible combination of inputs that include internal temperature and 4 universal inputs each which means they can be individually configured with the software to accept any combination of the following input types: 

RTD / Pt100
Thermocouple (Types J, K, T, R, S, B, N) with isolated tip
4/20mA (Scaled in Engineering Units)
0-10VDC (Scaled in Engineering Units)
±100mV
Various Thermistors
Slidewire
Cu Sensors
Ni Sensors

application examples

  • Baking
  • Flow Monitoring
  • Fridge & Freezer Monitoring
  • Cryogenic
  • Weather Environment
  • Process Control
  • Foundry
  • Flow Meter Recording

specification

  • Power : Internal Battery - LAA (3.6V DC) 2.2 Ah user replaceable Lithium-Thionyl-Chloride 
  • Battery Life : Depending on input configuration, >5 years with a sample rate > 2 minutes and LED flash rate > 30 second
  • Ambient Temperature : -30°C to +70°C (-22°F to 158°F)
  • Number of Readings Stored : 128,000 Readings
  • Memory Utilisation : Wrap Around (First in, First Out) or Stop when Full (Default).
  • Sample Rate  : 1 Second to 24 hours for all channels
  • Sample Method :Point - The sampled measurement is stored directly to memory. Averaging - The average value per period is stored, Maximum : The maximum value per period is stored Minimum - The maximum value per period is stored
  • Reusable When Memory Full : Yes
  • Software Alarms : Yes, high and low alarms can be specified in the software
  • Intrinsic Safety : No
  • Channel Indicators : 8 Green LED's to indicate logging status changing to Red to indicate alarm conditions.
  • Communication Interface : USB Isolated (mini-USB 1B Male connector) - Cable Supplied
  • Material : ABS - Moulded from RoHS compliant material
  • Dimensions : 124mm (L) x 78mm (W) x 32mm (H)
  • Weight : 165 grams
  • UL Flame Rating
    • 94-VO at 0.059' wall
    • 94-5VA at 0.098' wall
    • UL File Number: E56070
  • Warranty - 3 Year (excluding battery)

Measuring Ranges

  • Internal Temperature : -30°C to +70°C (-22°F to +158°F)
  • Internal Humidity : 0 to 100%RH
  • Plug In External Temperature Sensor : -30°C to +70°C (-22°F to +158°F)
  • Plug In External Humidity Sensor : 0 to 100%RH
  • RTD / Pt100 to EN 60751 : -200°C to +700°C (-328°F to +1292°F)
  • RTD (Pt100) JISC : -100°C to +457°C (-148°F to +854°F)
  • Thermocouple Type K : -200°C to +1370°C (-328°F to +2498°F)
  • Thermocouple Type J : -200°C to +1200°C (-328°F to +2192°F)
  • Thermocouple Type T : -200°C to +400°C (-328°F to +752°F)
  • Thermocouple Type S : 100°C to +1760°C (+212°F to +3200°F)
  • Thermocouple Type R : 100°C to +1760°C (+212°F to +3200°F)
  • Thermocouple Type B : +700°C to 1820°C (+1292°F to +3308°F)
  • Thermocouple Type N : -180°C to +1300°C (-292°F to +2372°F)
  • Ni120 : -80°C to +230°C (-112°F to +446°F)
  • Ni100 : -60°C to +180°C (-76°F to +356°F)
  • Cu10 : -200°C to +260°C (-328°F to +500°F)
  • Cu100 : -100°C to +260°C (-148°F to +500°F)
  • Thermistor (TH-001) : -40°C to +140°C (-40°F to +284°F)\
  • Slidewire : 0 to 100%
  • MilliVolt : -110mV to +110mV (Over Voltage protected to 24VDC) scaled in engineering units
  • 0-10VDC : 0VDC to 11VDC (Over Voltage protected to 24VDC) scaled in engineering units          
  • 0-1VDC : 0VDC to 1.1VDC (Over Voltage protected to 24VDC) scaled in engineering units
  • 4/20mA : 0.1 to 21mA scaled in engineering units
  • Pulse Counter 
    • Maximum Count = 65535 per sample period.
    • Maximum Frequency = 32678Hz.
    • Logic 0 < 0.9Vdc.
    • Logic 1 > 1.9Vdc
    • Trigger: Negative Edge
    • User settable filter link 32768Hz or 50Hz. (default: 32768Hz)
    • Can be used with common collector outputs when using an external shunt resistor.

Measuring Accuracies

  • Internal Temperature : ± 0.5°C (-30°C to +70°C)
  • Internal Humidity : ± 2.0%RH over 10%-90%, ± 4.0% RH over the range 0-10% and 90-100%RH
  • Plug In External Temperature Sensor : ± 0.3°C @ 25°C, ± 0.5°C over the range 0°C to +50°C, ± 1.0°C over the range -30°C to 0°C and 50°C to 70°C
  • Plug In External Humidity Sensor : ± 2.0%RH over 10%-90%, ± 4.0% RH over the range 0-10% and 90-100%RH
  • RTD / Pt100 to EN 60751 : ± 0.1°C + 0.1% rdg
  • RTD (Pt100) JISC : ± 0.1°C + 0.1% rdg
  • Thermocouple Type K : ± 1.9°C
  • Thermocouple Type J : ± 1.7°C
  • Thermocouple Type T : ± 0.9°C
  • Thermocouple Type S : ± 2.3°C
  • Thermocouple Type R : ± 2.3°C
  • Thermocouple Type B : ± 2.4°C
  • Thermocouple Type N : ± 1.8°C
  • Ni120 : ± 0.2°C ± 0.1% rdg
  • Ni100 : ± 0.2°C ± 0.1% rdg
  • Cu10 : ± 3°C ± 0.5% rdg
  • Cu100 : ± 0.2°C ± 0.1% rdg
  • Thermistor (TH-001) : Type dependent, please contact sales office for more details
  • Slidewire : ± 0.02% FSD
  • MilliVolt : ± 0.01% FSD
  • 0-10VDC : ± 0.01% FSD       
  • 0-1VDC : ± 0.01% FSD
  • 4/20mA : 
    • ± 0.01% FSD (for Isolated Loops), accuracy may be reduced for non-isolated loops.
    • Some applications may require the use of an external isolator (not supplied).
    • Maximum Volt drop in a 4/20mA loop = 4VDC
    • Will also accept 0/20mA signals
  • Pulse Counter : 20ppm

To view the full specification please download the datasheet below.

description

RTD / Pt100
Thermocouple (Types J, K, T, R, S, B, N) with isolated tip
4/20mA
0-10VDC
±100mV
Various Thermistors
Slidewire
Cu Sensors
Ni Sensors

The SL7104 data logger also has digital input which can be used as a trigger to start logging or as a pulse counter for interfacing with flow meters or other sensors with a pulse output.

All SL7104 data loggers are supplied with a single computer licence of our TempIT-PRO software. Saving £157.00 

Enquire About This Product